News

 
  1. RTI Electronics AB is Looking for New Employees

    26 Jan 12 Would you like to be part of developing the medical technology products and markets of the future? RTI Electronics is growing and is now looking for new co-workers who want to be part of shaping the future in an exciting industry.

  2. RTI Electronics, Inc. has Moved

    02 Jan 12 RTI Electronics, Inc. has moved to larger location with lots of space to grow. Starting Tuesday, January 3, 2012 you can find RTI Electronics, Inc. at:

    33 Jacksonville Road
    Building 1
    Towaco, NJ 07082

    RTI Inc's phone numbers and contact information are the same for your convenience.

  3. Merry Christmas 2011 from RTI

    All the Best for 2012

    22 Dec 11 We wish to thank our customers and users for a fruitful 2011, wish you a really nice Christmas, and the best for 2012. Monday 26 December RTI Headquarters is closed.

Show all news

Solidose 300

Solidose 300

Simplifies basic X-ray measurements

Solidose is RTI's user-friendly dosimeter and rate meter that is used for the quality control and servicing of X-ray equipment. The Solidose 300 contains built-in applications so that measurements can be performed as easily as possible. You simply select the module and the calibration application that meet your needs. The Solidose 300 is supplied with two ADI modules (automatic detector identification) and more can be purchased afterwards.

Reduced Solidose Price

Right now, we offer discounted prices on our popular Solidose meters. Contact your local distributor to get more information or to place an order.


Solidose 300 comes with
  • R100 Solid state detector
  • WinRAD Software
  • Calibration application 8/B For standard radiogrpahy. Reference point at 70 kV/3 mm Al.
  • Calibration application 8/D For skin dose mammography. Reference point at 28 kV/30 µm Mo.
Details

  • A single measuring range covering more than six decades (>120 dB).
  • Built-in applications such as HVL and reproducibility.
  • Automatic Detector Identification (ADI) system, fully compatible with PMX-III .
  • RS-232 computer interface with software (WinRAD).
  • Battery operated or external power supply.
  • Possible to connect other solid state detectors from RTI, using ADI system. 
  • Use of solid state detectors and other probes.

Main Features

Dose and dose rate
Radiography
CT
Mammography
Input dose rate to image intensifier
Reproducibility
HVL
Relative mAs
Dose/frame
Ratio

Optional calibrations (ADI)

 

  • 8/A For heavy filtered beams, for example input dose to the I.I. Reference point at 70 kV/23 mm Al.
  • 8/C For film dose mammography. Reference point at 28 kV/2 mm Al + 30 µm Mo.

 

 

Specifications
Charge  Range  0.5 pC-10 mC
Current  Range  5 pA-25µA
Inaccuracy ± 1% or ±0.5 pA
Dose (R100)  Range  10 nGy-180 Gy
1.15 µR-23 000 R
Inaccuracy  ±5 %
Dose rate (R100)   Range 100 nGy/s-450 mGy/s
11.5 µR/s-50 R/s
Inaccuracy ±5 %
Random error  ±1 % 
Display  Alpha-numerical 16x2 LCD
Connector  Trial LEMO type "øs"
Computer interface  RS-232
Dimensions  205x135x58 mm
Weight  1.1 kg
Power source  4 power alkaline batteries type LR6 (size AA)
Battery life  20 hours
Power supply  12 v AC/DC adapter (optional)

 

 
 

Please enter your user name and password below.

If you are the proud owner of a Barracuda, Piranha, or Nova you just have to enter your serial number as the user name and leave the password box empty.