News

 
  1. Public Holiday and Changed Opening Hours

    16 May 12 Between 17–18 May RTI Headquarter is closed.
    RTI Inc is open as usual.

  2. New Software Release – 2012.05A

    11 May 12 Piranha and Barracuda Latest Update

    Released: May 11, 2012

    Ocean: 2012.05.09.82
    QABrowser: 4.2B
    oRTIgo: 6.5A
    CT Dose Profile Analyzer: 4.1A
    Firmware: 2012.2A

  3. Mammo12

    One Shot HVL for Mammography

    07 May 12 One shot HVL for mammography is now available to all new Piranhas of version 2 or higher.
    One shot HVL is possible both if you are using Ocean or QA Browser, with or without compression paddle.

    All Piranhas (version 2 or higher), which are in for re-calibration, will get the one shot HVL feature.

  4. Time for Calibration - Schedule Yours Now

    06 May 12 Please take five minutes to schedule your coming calibration with the RTI Service department.

Show all news

Piranha Panoramic Holder (OPG)

Piranha Panoramic Holder (OPG)

Field Position Analyzer that Saves Time and Money

The Piranha is ideal for measurement on panoramic (OPG) systems. The small detector area and the "Piranha position check", a feature that makes it possible to verify the detector position, ensures more accurate and more reproducible results than with other devices, since the Piranha works wireless, you do not have to worry about any cable during the rotation of the OPG. The Piranha OPG holder makes it possible to attach the Piranha on the panoramic system and quickly perform the measurement. Of course, the Piranha is perfectly suited for measurements on standard intra oral systems as well.


The Piranha Panoramic Holder can be fitted with magnets or adhesive tape. The holder comes with several bendable plates that can be used to fit "around corners" depending on the geometry of the panoramic (OPG ) system. Two screws makes it possible to adjust the horizontal position +/-2 mm.

This means that the holder doesn't need to be positioned "perfectly" with the magnets or tape but can be adjusted afterwards. Note that on some panoramic systems, magnets are not allowed or not possible to use, and the plates makes it easy to use adhesive tape instead.

 

To perform a measurement:
  1. Attach the holder and MPD.
  2. Select the "MPD Position Check" and make an exposure.
  3. Select "measure all".
  4. Make an exposure and get kVp, exposure time, dose, dose rate, HVL total filtration, and waveform. All in one exposure!

Article Nr: 9744001-00

 
 