News

 
  1. Public Holiday and Changed Opening Hours

    16 May 12 Between 17–18 May RTI Headquarter is closed.
    RTI Inc is open as usual.

  2. New Software Release – 2012.05A

    11 May 12 Piranha and Barracuda Latest Update

    Released: May 11, 2012

    Ocean: 2012.05.09.82
    QABrowser: 4.2B
    oRTIgo: 6.5A
    CT Dose Profile Analyzer: 4.1A
    Firmware: 2012.2A

  3. Mammo12

    One Shot HVL for Mammography

    07 May 12 One shot HVL for mammography is now available to all new Piranhas of version 2 or higher.
    One shot HVL is possible both if you are using Ocean or QA Browser, with or without compression paddle.

    All Piranhas (version 2 or higher), which are in for re-calibration, will get the one shot HVL feature.

  4. Time for Calibration - Schedule Yours Now

    06 May 12 Please take five minutes to schedule your coming calibration with the RTI Service department.

Show all news
Chamber Adapter

Magna 1 cc

Magna 1 cc

Designed Especially for Mammographic Measurements

The Magna 1cc ionisation chamber is designed especially for mammographic dose and dose rate measurements. However, it has an excellent energy response and can also be used for radiographic applications. The Magna Ion Chamber is the ideal choice for two reasons;
Evaluations of Magna chambers show an average response is within ± 1% over an extended mammography range of 20-40 kVp (HVL of 0.15 mm to 0.73 mm Al).
All mammographic mean glandular dose tables are based on in air measurements. MAGNA’s air equivalent construction makes it ideally suited for in air exposure measurements of mean glandular dose. The Magna 1 cc will be delivered in a durable case for its best protection.

Magna 1cc and Piranha

The Magna 1cc can be connected via Chamber Adapter to the Piranha to measure Mammo applications.

Magna 1cc together with Barracuda

The Magna 1cc can be connected directly to the Barracuda (via a bias electrometer module EMM-Bias module). It can be used especially for mammography applications when an ion chamber is preferred instead of a solid-state detector.

Magna 1cc together with Solidose 400

Magna 1cc can also be connected directly to the Solidose 400 - no adapter cable is needed. It can be used especially for mammography applications when an ion chamber is preferred instead of a solid-state detector.


General
  • Energy dependence Within ±1 % for HVL of 0.15 mm to 0,73 mm Al
  • Within ±2 % for 20 to 150 kVp
  • Construction Air equivalent conductive plastic
  • Sensitivity 38 nC/Gy or 0.33 nC/R
  • Collecting Volume 1 cm3
  • Ventage Yes
  • Leakage < 50 fA
  • Typical Bias 300 V DC
  • Window Kapton conductive film + mylar protective film
  • Cable length 2 m
  • Connector Triax LEMO
Ranges with Piranha (Chamber Adapter required)
Dose Rate 0.25 mGy/s to 2.5 Gy/s 
Inaccuracy ±6 % or ±0.025 mGy/s

Ranges with Barracuda (with EMM-Biasw)
Dose  150 nGy-at least 2 MGy ±5 %
17 µR-200 MR
Dose rate  5 µGy/s-250 Gy/s ±5 %
30 MR/min-1500 kR/min
Ranges with Solidose 400
Dose 13 µGy -200 kGy ±5 %
1.5 mR-20 MR
Dose rate  125 µGy/s-390 Gy/s ±5 %
0.85 R/min-2.6 MR/min

Article Nr: 9706100-00

 
 