The MAS-2 Probe is a current clamp probe for non-invasive measurement of mA (tube current) and mAs. The greatest advantages are that the probe is easy to use and that it is safe. Furthermore there is no need to disconnect the high voltage cables to be able to measure mA and mAs. The probe is simply clamped on to the high voltage cable and then ready for measurements, no connection inside the X-ray generator is required. Together with your meter direct reading of mA and mAs as well as waveform are obtained.
A special model, MAS-2B optimized for the Barracuda is available. The MAS-2B Probe is ideal for QA since no connection inside the X-ray generator is required and it allows anyone to measure mA and mAs. The Barracuda MAS-2 Probe can be used to measure tube current down to 10 mA. To use the Barracuda MAS-2 Probe the Barracuda must have an electrometer module. Recommended model is EMM-BiasW. The Barracuda MAS-2 probe can be used with the extension cable, EXT-1. Used with the Barracuda, the MAS-2 provides a non-invasive way to measure tube current and mAs on X-ray equipment.
Connected to the Piranha, the MAS-2 is ready for non-invasive measurements, reading mA, mAs, and capturing an mA waveform. The Piranha MAS-2 Probe is ideal for QA since no connection inside the X-ray generator is required and it allows anyone to measure mA and mAs. The Piranha MAS-2 Probe can be used to measure tube current down to 10 mA.
The MAS-2 probe supports the Piranha ADI (automatic detector identification) system. All information regarding the Probe is stored in a memory inside the probe. When connected, the Piranha automatically identifies the probe and makes all necessary adjustments without any need for interaction from the user. The Piranha ADI system also gives full interchangeability of probes between different Piranha systems using the same type of probes.
MAS-2 is a non-invasive current clamp probe. It provides a non-invasive way to measure mA and mAs on X-ray equipments. If the same MAS-2 probe should be used with both Barracuda and Solidose 300/400, you must choose this model of the MAS-2 probe.
| Type | MAS-2B |
| Current range | 10-4000 mA |
| Inaccuracy | ±5 % or ±2 mA (±3 % at 250 mA) |
| Charge range | 0.1 mAs-9999 mAs |
| Frequency range | DC-500 Hz (-3 dB) |
| Size | 183 x 61 x 36 mm |
| Connector | LEMO Triaxial |
| Type | MAS-2 |
| Current range | 10 - 4000 mA |
| Inaccuracy | ±5 % or ±2 mA (±3 % at 250 mA) |
| Charge range | 0.1 mAs - 9999 mAs |
| Frequency range | DC - 500 Hz (-3 dB) |
| Size | 183 x 61 x 36 mm |
| Connector | Hirose with built-in ADI (automatic detector identification) system |
| mAs | 1.0 mAs - |
| mA | 10 mA - 4000 mA |
| Scale factor | 1 nA/mA |
| Size | 183 x 61 x 36 mm |
| Frequency range | DC - 5/500 Hz (-3 dB) |
| Maximum cable diameter | 24 mm |
| Connector | LEMO triax |
| Power source | 9 V "6LR61" Alkaline battery |
| Measuring method | Non-invasive |
| Inaccuracy | ±5 % |