News

 
  1. assets/News/Juli-2010/_resampled/SetWidth190-South-2.jpg

    RTI Electronics Strengthening its Position in South America

    23 Jul 10 We are now pleased to present three new distributors in South America.

  2. assets/News/Juli-2010/_resampled/SetWidth190-Picture-T20-034.jpg

    Important Solid State Detector Accessories by RTI Electronics

    23 Jul 10 If you own an RTI Piranha or Barracuda multi-meter, you may want to consider adding two special solid state detectors specially developed by RTI Electronics.

  3. Exciting Innovations at AAPM

    14 Jul 10 RTI's display 756 will feature some of the most exciting X-ray QA products.

  4. assets/About_Us/Certification-and-Approval/_resampled/SetWidth190-Accred-2021.png

    Quality-Assured Laboratories at RTI Electronics AB

    14 Jun 10 This accreditation is a guarantee that our assignment is carried out impartially, fairly and based on internationally accepted standards. It provides third-party assurance that we have the quality management system to ensure all of our calibrations to be consistently of the highest quality.

  5. assets/Accessories/MAS-1-o-2/_resampled/SetWidth190-MAS-1.png

    Redesigned and Improved MAS-1

    02 Jun 10 This month RTI releases the new redesigned MAS-1.

Show all news
MAS-1 Probe

MAS-2 Probe

MAS-2 Probe

Simple Way for Non-invasive Measurements

The MAS-2 Probe is a current clamp probe for non-invasive measurement of mA (tube current) and mAs. The greatest advantages are that the probe is easy to use and that it is safe. Furthermore there is no need to disconnect the high voltage cables to be able to measure mA and mAs. The probe is simply clamped on to the high voltage cable and then ready for measurements, no connection inside the X-ray generator is required.  Together with your meter direct reading of mA and mAs as well as waveform are obtained.


MAS-2 Probe together with Barracuda

A special model, MAS-2B optimized for the Barracuda is available. The MAS-2B Probe is ideal for QA since no connection inside the X-ray generator is required and it allows anyone to measure mA and mAs. The Barracuda MAS-2 Probe can be used to measure tube current down to 10 mA. To use the Barracuda MAS-2 Probe the Barracuda must have an electrometer module. Recommended model is EMM-BiasW. The Barracuda MAS-2 probe can be used with the extension cable, EXT-1. Used with the Barracuda, the MAS-2 provides a non-invasive way to measure tube current and mAs on X-ray equipment.

MAS-2 Probe together with Piranha

Connected to the Piranha, the MAS-2 is ready for non-invasive measurements, reading mA, mAs, and capturing an mA waveform. The Piranha MAS-2 Probe is ideal for QA since no connection inside the X-ray generator is required and it allows anyone to measure mA and mAs. The Piranha MAS-2 Probe can be used to measure tube current down to 10 mA.

The MAS-2 probe supports the Piranha ADI (automatic detector identification) system. All information regarding the Probe is stored in a memory inside the probe. When connected, the Piranha automatically identifies the probe and makes all necessary adjustments without any need for interaction from the user. The Piranha ADI system also gives full interchangeability of probes between different Piranha systems using the same type of probes.

MAS-2 Probe together with Solidose 300 or 400

MAS-2 is a non-invasive current clamp probe. It provides a non-invasive way to measure mA and mAs on X-ray equipments. If the same MAS-2 probe should be used with both Barracuda and Solidose 300/400, you must choose this model of the MAS-2 probe.

Specifications with Barracuda
Type  MAS-2B
Current range                         10-4000 mA
Inaccuracy  ±5 % or ±2 mA (±3 % at 250 mA)
Charge range 0.1 mAs-9999 mAs
Frequency range DC-500 Hz (-3 dB)
Size 183 x 61 x 36 mm
Connector LEMO Triaxial
Specifications with Piranha
Type MAS-2
Current range 10 - 4000 mA
Inaccuracy ±5 % or ±2 mA (±3 % at 250 mA)
Charge range 0.1 mAs - 9999 mAs
Frequency range DC - 500 Hz (-3 dB)
Size 183 x 61 x 36 mm
Connector Hirose with built-in ADI (automatic detector identification) system
Ranges with Solidose 300/400
mAs              1.0 mAs -
mA             10 mA - 4000 mA
Specifications General
Scale factor  1 nA/mA
Size  183 x 61 x 36 mm
Frequency range  DC - 5/500 Hz (-3 dB)
Maximum cable diameter 24 mm
Connector  LEMO triax
Power source  9 V "6LR61" Alkaline battery
Measuring method  Non-invasive
Inaccuracy  ±5 % 
 
 

Please enter your user name and password below.

If you’re the proud owner of a Barracuda, Piranha, or Nova you just have to enter your serial number as the user name and leave the password box empty.