News

 
  1. assets/News/Juli-2010/_resampled/SetWidth190-South-2.jpg

    RTI Electronics Strengthening its Position in South America

    23 Jul 10 We are now pleased to present three new distributors in South America.

  2. assets/News/Juli-2010/_resampled/SetWidth190-Picture-T20-034.jpg

    Important Solid State Detector Accessories by RTI Electronics

    23 Jul 10 If you own an RTI Piranha or Barracuda multi-meter, you may want to consider adding two special solid state detectors specially developed by RTI Electronics.

  3. Exciting Innovations at AAPM

    14 Jul 10 RTI's display 756 will feature some of the most exciting X-ray QA products.

  4. assets/About_Us/Certification-and-Approval/_resampled/SetWidth190-Accred-2021.png

    Quality-Assured Laboratories at RTI Electronics AB

    14 Jun 10 This accreditation is a guarantee that our assignment is carried out impartially, fairly and based on internationally accepted standards. It provides third-party assurance that we have the quality management system to ensure all of our calibrations to be consistently of the highest quality.

  5. assets/Accessories/MAS-1-o-2/_resampled/SetWidth190-MAS-1.png

    Redesigned and Improved MAS-1

    02 Jun 10 This month RTI releases the new redesigned MAS-1.

Show all news

DCT10 Pencil Ionization Chamber

DCT10 Pencil Ionization Chamber

Ionization Chamber for CTDI Measurements

The DCT10 RS is rugged, cylindrical pencil shaped air ionization chamber for CTDI measurements. The CT probe is intended for measuring and monitoring the exposure output level of CT scanners in a phantom or in air. CT dose index, CTDI, can be measured in accordance with IEC 61223-2-6.


DCT10 and Barracuda

The DCT10 RS can be connected directly to the Barracuda (via an electrometer module supporting bias) to measure CT dose. Barracuda has built-in functions to simplify measurements of CTDI (CT dose index).

Measuring ranges with Barracuda (with EMM-Biasw)
Dose 40 nGym-350 kGym
4 µGycm-35 MGycm

Dose rate 1.6 µGym/s-70 kGym/s
160 µGycm/s-7kGycm/s
Inaccuracy ±5 % or 16 µGycm/s

DCT10 and the Solidose Family

The DCT10 RS can be connected directly to the Solidose 400 to measure CT dose.

Measuring ranges Solidose
Dose 20 µGycm - 400 kGycm ±5 %
2.3 mRcm - 46 MRcm
Dose rate 200 µGycm/s - 600 Gycm/s ±5 %
1.4 Rcm/min - 4 MRcm/min

Specifications

Connector type LEMO triaxial or
BNC banana
Cable 2 m, low noise triaxial
Active volume 4.9 cm3
Active length 100 mm
Inner diameter 8.0 mm
Typical leakage ±4 fA
Radiation quality 80 - 150 kV
Sensitivity 26 nC/Gycm, 260 nC/Gy
0.5 Rcm/min - 2700 kRcm/min

 
 